Hello, welcome to Jiangsu Huaqe Electric Co., Ltd.!

华齐产品
Your current location : home >> Products >> Online analysis instrument

YLG-2058 Chinese Online Residual Chlorine Analyzer

2020-06-06 13:52:08
second
YLG-2058 Chinese Online Residual Chlorine Analyzer
Detailed introduction:

YLG-2058 Chinese Online Residual Chlorine Analyzer

●Overview

  YLG-2058 Chinese online residual chlorine analyzer is a brand new residual chlorine analyzer of RAMZY company. This instrument is a highly intelligent online monitor. It consists of a secondary meter a sensor, a three-component plexiglass flow cell. Can simultaneously measure residual chlorine, pH value, temperature. It can be widely used for continuous monitoring of residual chlorine pH value of various water qualities in electric power, waterworks, hospitals other industries.

●Features

● Chinese display, Chinese menu operation: the operation is simple, the operation steps are prompted in Chinese throughout the process, convenient fast

● High intelligence: YLG-2058 Chinese online residual chlorine analyzer adopts high-precision AD conversion single-chip micro-processing technology, which can complete residual chlorine, pH temperature measurement, instrument self-test other functions

● Multi-parameter display on the same screen: display residual chlorine, temperature, pH value, output current, status time on the same screen

● Current isolation output: adopts photoelectric coupling isolation technology, which has strong anti-interference ability can be transmitted remotely

● High low alarm function: High low alarm isolation output, hysteresis can be adjusted arbitrarily

● Waterproof dustproof design: Protection grade IP65, suitable for outdoor use

●Technical parameters

Measuring range: residual chlorine: 0~20.00mg/L, resolution 0.01mg/L

                 HOCL: 0~10.00mg/L, resolution 0.01mg/L

                 pH value: 0~14.00pH, resolution 0.01pH

                 Temperature: 0~99.9℃, resolution 0.1℃

Precision: residual chlorine: ±2% ±0.035mg/L, whichever is greater

HOCL: ±2% ±0.035mg/L, whichever is greater

pH value: ±0.05pH; temperature: ±0.5℃ (0~60.0℃)

Sample temperature: 0~60.0℃, 0.6MPa

Sample flow rate: 200 ~ 250 mL/1 minute, automatically adjustable

Minimum detection limit: 0.01mg/L

Galvanically isolated output: 0~10mA (load<1.5 kΩ), 4~20 mA(load<750Ω)

High low alarm relay: AC220V, 7A; any adjustment of hysteresis 0~5.00mg/L

RS485 communication interface (optional) can be easily connected to a computer for monitoring communication

Number of data storage: 1 month (1 point/5 minutes)

Power source: AC220V±22V, 50Hz±1Hz; optional DC24V power supply

Protection level: IP65

Dimensions: 146 (length) × 146 (width) × 108 (depth) mm

Hole size: 138×138 mm, Note: It can also be installed on the wall, please specify when ordering

Weight: Secondary table: 0.8kg, residual chlorine in the flow cell, pH electrode weight 2.5kg

Working conditions: ambient temperature 0~60℃ relative humidity <85%

Use flow-through installation, inlet outlet diameter Φ10

Note: Two current outputs are optional. One residual chlorine corresponds to the current output the other optional pH temperature corresponds to the current output. The factory standard configuration is one-channel residual chlorine corresponding current output. Please specify when ordering.